SCOPE OF CALIBRATION FACILITY (Electro-Technical Calibration)
Quantity Measured/ Instrument |
Range / Frequency | *Calibration Measurement Capability (+ -) | Remarks |
SOURCES (at Lab & Site) | |||
1. DC Voltage | 1 mV to 300 mV 300mV to 30 V 30 V to 300 V 300 V to 1000 V |
1.178% to 0.019% |
Using Fluke 5080 A 240 |
2. DC Current | 100 μA to 1 mA |
0.202% to 0.141% |
Using Fluke 5080 A 240 with Current Coil 5500 A By Direct Method |
3. AC Voltage | 50 Hz to 1 kHz 10 mV to 30 mV 30 mV to 30 V 30 V to 300 V 300 V to 1000 V |
1.075% to 0.613% |
Using Fluke 5080 A 240 By Direct Method |
4. AC Current |
50 Hz to 1 kHz 30 μA to 300 μA 300uA to 3.0 mA 3.0 mA to 10 A 20 A to 1000 A |
|
Using Fluke 5080 A 240 with Current Coil 5500 A By Direct Method |
5. Resistance (Discrete Values) (2 Wire & 4 Wire) |
1 Ω, 10 Ω, 19 Ω, 100 Ω,
|
1.362%,0.189%,0.122%,0.047%,
|
Using Fluke 5080 A 240
DIRECT SHUNT RESISTORS |
6.Frequency | 45 Hz to 1 kHz | 0.017% to 0.058% |
Using Fluke 5080 A MFC By Direct Method |
7. STOP WATCH/TIMER/ HOUR METER – time interval meter# | Upto 10 s 10 s to 100s 100s to 999 s |
0.0001 s 0.02 s 0.2s |
Using Digital Time Calibrator Make: Ajit by Direct Method |
8. Power Factor Meter | 0.1 Lead/Lag (U.P.F.). | 0.1 P.F. | Using Fluke 5080 A MFC By Direct Method |
9. Power (U.P.F.) 1 φ |
100 mV to 1000 V 0.5 A to 20 A 0.5 W to 20 kW |
1% | Using Fluke 5080 A MFC By Direct Method |
* Measurement Capability is expressed as an uncertainty (+) at a confidence probability of 95%
$Only in Permanent Laboratory
♣Only for Site Calibration
# The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual environmental conditions and master equipment used.
SCOPE OF CALIBRATION FACILITY (Electro-Technical Calibration)
Quantity Measured/ Instrument |
Range / Frequency | *Calibration Measurement Capability (+ -) | Remarks |
MEASURE (At Lab & Site) | |||
1. DC VOLTAGE# | 1 mV to 100 mV 100 mV to 10 V 10 V to 1000 V |
0.052% to 0.005% |
Using Fluke 8846 A 6½ DMM |
2. AC VOLTAGE# | 50 Hz to 1 kHz
|
0.160% to 0.141%
|
Using Fluke 8846 A 6½ DMM By Comparison/ Direct Method |
3. DC CURRENT# | 10 μA to 100 μA 1mA to 10 mA 100mA to 390 mA 390 mA to 10 A |
1.28% to 1.273% |
Using Fluke 8846 A 6½ DMM By Comparison/ Direct Method |
4. AC CURRENT # |
50 Hz to 1 kHz 100 μA 100 μA to 1 mA 10 mA to 100m A 390m A to 10 A |
|
Using Fluke 8846 A 6½ DMM By Comparison/ Direct Method |
5. AC HIGH CURRENT $ | 10 A to 5000A | +0.5%RDG, |
Using Fluke 8846 A 6½ DMM & Standard CTs by Direct Method/comparison method |
6.FREQUENCY# | 45 Hz to 400 Hz | Less than 1% |
Using 5½ digit universal calibrator by Direct Method |
7. DC RESISTANCE# (2 Wire & 4 Wire) |
0 to 1 Ω 1 Ω to 100 Ω 100 Ω to 1 kΩ 10 kΩ to 100kΩ 1M Ω to 10M Ω 100M Ω to 1000M Ω |
0.048% 0.048% to 0.012% 0.012% to 0.013% 0.012% to 0.012% 0.013% to 0.047% 0.925% |
Using Fluke 8846 A 6½ DMM By Comparison/ Direct Method |
8. Power Energy 1 φ & 3 φ |
50 Hz 230 V to 300 V 10 A to 120 A 0.5 P.F. to U.P.F. |
0.5% | Using power analyzer KI 5 |
9. RESISTANCE# | 10 MΩ 100 MΩ 10 MΩ to 1 GΩ 1 GΩ to 10 GΩ |
2% | Using Standard Decade Meg Ohm Box |
10. STOP WATCH/TIMER/ HOUR METER – time interval meter# | 0.001mSec to 10 s |
0.0001 s |
Using Digital Time Calibrator Make: Ajit |
11. DC HIGH VOLTAGE♣ (Source-#) |
1.0 kV to 100 kV |
0.05 kV to 0.95 kV |
Using HV divider with DMM by Direct Method |
12. AC HIGH VOLTAGE♣ (Source-#) |
1.0 kV to 100 kV |
0.03 kV to 2 kV |
Using HV divider with DMM by Direct Method |
* Measurement Capability is expressed as an uncertainty (+) at a confidence probability of 95%
$Only in Permanent Laboratory
♣Only for Site Calibration
# The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual environmental conditions and master equipment used.